SS1000 - Portable Carbon Dioxide Analyzer Solution

SpectraSensors SS1000 Portable Carbon Dioxide Analyzer is light weight, easy to handle and battery-powered.  Utilizing the same extraordinary sensor technology as the SS2000, the instrument is designed for convenience and allows for fast and easy measurements in the field.
 
Product Code: 10603



Application Note

SpectraSensors SS1000 Portable is a light weight, easy to handle, battery-powered CO2 analyzer, used to verify measurements and for spotchecking when other methods provide questionable results.
 
In natural gas pipeline applications, poor quality measurement results are extremely costly. Additional processing or dehydration costs, upset conditions, shut-ins and inconsistent process results may be caused by sensors that do not perform properly. The SS1000 reveals poorly performing sensors, pinpoints high moisture and can be used as a standard for measurement validation.
 
Rapid Response Time The SS1000 allows for fast, simple operation. The analyzer’s laser and detector take measurements 4 times per second and average the results. These real-time measurements are not hampered by wetup (absorption) or dry-down desorption) as with surfaced-based sensors.
 
Reliable Using state-of-the-art laser technology developed by NASA, the SS1000 analyzer is more reliable, with greater repeatability than surface based sensors and is not subject to the interpretation errors of a chilled mirror.
 
No Interference As with all SpectraSensors analyzers, the SS1000 combines a patented laser with control electronics and “smart” software. The analyzer’s sensor heads are not subjected to corrosives or contaminants in the gas because the sensor is isolated from the sample gas stream.
 
The result is an analyzer which does not suffer from contamination or drift due to vapor impurities such as glycol, methanol, amines, hydrogen sulfide, or mercaptans.
 

Payback The SS1000 very quickly pays for itself eliminating the cost of consumables, extra sensor heads, factory calibrations and labor and overhead associated with excessive maintenance. Expensive problems caused by unreliable gas measurements such as added processing steps and poor gas quality can be eliminated.

 

« Previous | Next »

Data Sheet


Performance

Target Components

CO2 in Natural Gas

Typical Measurement Ranges

0-10%*

Typical Precision

±1% of reading or ±0.04ppmv (400ppmv)
whichever is greater

Measurement Response Time

0.25-2 sec

(Total system response is dependent on flow rate
and sample system volume)

Principle of Measurement

Tunable Diode Laser Absorption Spectroscopy

Environmental Temperature Range

-20º to 50ºC (-4º to 122ºF)

Sample Inlet Pressure

10 to 25 PSIA, 10 PSIG Maximum

(70-170kPa Abs, 70kPaG maximum)

Sample Cell Temperature Range

-20º to 50ºC (-4º to 122ºF)
-15º to 60ºC (-5º to 140ºF)

Maximum Cell Pressure

70 kPag (10 PSIg)

Sample Flow Rate
0.1-10 L/min (60.2 to 20 scfh)
Recommended Validation Binary Cal Gas Bottle with Methane Background
 

Electrical

Power

100-240 VAC, 50-60 Hz standard
12-Volt, Sealed Lead-Acid Battery
Approx. 8 hours usage time per charge

Max Current

0.5A @ 120 VAC during recharging

Controller to Cell Cable Length

0.5m standard

Communication

RS232C - all parameters
LCD Display
Concentration, Cell Pressure and Temperature & Diagnostics
 

Physical

Dimensions

Nominal 200mm H x 175mm W x 450mm H
(8" H x 7" W x 18" D)

Weight Approximately

Approx 6.8Kg (15lbs)

Sample Cell Dimensions

438mm H x 108mm W
(17-1/4” H × 4-1/4” W)

Sample Cell Construction

316L Series Polished Stainless Steel standard

Number of Sample Cells
1
Area Classification
Non-Hazardous (certified) locations - General Purpose


 

Drawings

 
SS1000 Portable Analyzer

 
 
 
 

« Previous | Next »

Typical Accessories


02199-00001
Spares Kit- includes fuses, o-rings and hardware
 
00000-02015 Extended calibration range (up to 100lb/MMSCF H2O)
Other ranges available (consult factory)
 
39901-12012
Replacement battery charger
 
01902-93001
Instruction Manuals (up to 2 copies at no charge)
 
00403-90001
Quick-connect flow meter
 

Replacement mating quick-connect fittings
(1 ea. input and output)
 
01910-30017
Stainless Steel Tag, up to 2 lines of text
 
00000-02014 Certified As-Built Drawings
2 sets
 
 

Photos

 

« Previous | Next »

Request a Quote

To request a quote online, please enter the following and we'll get back to you promptly:

Your Name*
Company
Email*
Address*
Address Line 2
City*
State
Zip Code
Country*
Phone
Product Code
Application Name / Stream Details
Measurement Range
Comments and Questions

« Previous | Next »