SS2100 - LNG Product (Export/Import) Analyzer Solution

A new technology to monitor H2O concentration on Vaporized LNG Streams

Product Code: 44401
 
 
 
 
 

Application Note

LNG Vaporization
Liquefied Natural Gas (LNG) is produced and shipped at temperatures of approximately -160oC, so there is little likelihood of significant water concentrations in the LNG itself. However, there are a number of different approaches for the regasification of LNG, using air, gases or liquids heated by burning the gasified LNG and one of the most common is Open Rack Vaporizers using seawater as the heating medium.

Critical Control of Moisture
Monitoring of vaporized LNG for water may be critical in some processes as an indicator of leakage across the heat exchangers. In some cases, the gasified LNG may be mixed with other fuels that are wet.

Traditional Measurement Solutions

Electrochemical and vibrating quartz crystal moisture analyzers have been the traditional method for monitoring the levels of moisture in LNG Product.

Unfortunately, trace hydrocarbons and other contaminants present in the stream contaminate the electrochemical probes. Also the probe calibration tends to drift over the lifetime of the sensor resulting in high maintenance costs due to the need to constantly replace the sensor. It’s not uncommon for moisture probes to be replaced multiple times a year. Quartz crystal moisture meters have slow wet-up and dry-down times that may distort the true moisture content during an upset.

SpectraSensors' Solution

SpectraSensors SS2100 is the ideal solution for this challenging application. Its non-contact laser and detector is impervious to damage from contaminants. Tunable Diode Laser technology means that measurement interferences from other infrared absorbing compounds are avoided. There are no wet-up or dry-down delays, resulting in fast updates even when the concentration changes dramatically. Only SpectraSensors employs Differential Spectroscopy, which incorporates a dryer using metal getter technology to subtract the spectrum of the dry gas from the wet spectrum. This allows the SS2100 to measure streams whose background gas composition changes.

Validation
Validation of low level (<10 ppm) water measurements is essential, but also extremely difficult, as certified blends at such low levels are not readily available. Some have attempted to use permeation devises, but these require regular maintenance to prevent drying out, and there is no independent way to verify the accuracy of the permeation device. SpectraSensors employs a dilution system that dilutes a relatively high concentration certified standard of H2O in Methane using dried sample gas.
 

Typical Background Stream Composition for LNG

Component

Minimum (Mole%)

Normal (Mole%)

Maximum (Mole%)

Water

0.05 ppm

<1 ppm

10 ppm

Nitrogen

0

1 – 3

3

Oxygen

0

<1

1

Methane

90

90 - 100

100

Carbon Dioxide

0

0

0.001

Ethane

0

0 - 10

10

Propane

0

0 - 5

5

Butanes

0

0 - 2

2

C5+

0

0 – 0.5

0.5

 

The background stream composition must be specified for proper calibration and measurement performance. Specify the Normal composition, along with the minimum and maximum expected values for each component, especially water, the measured component. Other stream compositions may be allowable with approval from SpectraSensors. 

 
 
 

« Previous | Next »

Data Sheet


Performance

Target Components

H2O in LNG Product (export/import)

Typical Measurement Ranges

0 – 10 ppm* (other ranges available upon request)

Typical Precision

 ±0.05ppm or 2% of Full Scale

Measurement Response Time

1 to ~60 sec*

Principle of Measurement

Tunable Diode Laser Absorption Spectroscopy
Non-differential

Environmental Temperature Range

-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional

Sample Inlet Pressure

70 kPag (10 psig) typical

210 kPag (30 psig) maximum


Cell Temperature

Maintain at 50ºC +/- 0.2ºC

Maximum Cell Pressure

70 kPag (10 PSIg)

Sample Flow Rate
 
3 - 4 L/min (6.4 to 8.5 scfh)*
Validation
A certified blend of H2O in Methane balance
is diluted with dried sample in the sample conditioning
system under flow control
 

Electrical

Power

100-240 VAC, 50-60 Hz standard

Max Current

Controller: 1 A @ 120 VAC

Controller to Cell Cable Length

1m standard (3m, 5m & 10m available optionally)

Communication

Current Loop Output 4-20 mA Isolated, 1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
Digital Outputs
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm 
LCD Display
Concentration, Cell Pressure and Temperature, Diagnostic Data
 

Physical

Controller Enclosure

NEMA 4X - 304 Stainless Steel standard

Controller Dimensions

343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*

Weight Approximately

13.1 Kg (28.6 lbs)*

Sample Cell Dimensions

28m Herriot Cell
559 mm H × 127 mm W
(22” H × 5” W)

Sample Cell Construction

316L Series Polished Stainless Steel standard
Number of Sample Cells
1 (single channel SS2100) or
2 (dual channel SS2100)
Area Classification
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
 ll 2G Ex d llB+H2 T5; Tamb : -20 ÷ +60 °C
Dimensions with Sample System 1678 mm H x 613 mm W x 427 mm D
(66” H x 24-1/8” W x 16-13/16” D)
Weight with Sample System (approx.) 68 kg (150 lbs.)

* Application specific; consult factory

« Previous | Next »

Drawings

 
 

Typical Accessories


SS2100 ARM9 Electronics
 

80000-02199

Temperature Control Board

23000-00030

4-20 mA Current Loop Board
 

80000-02203

Power Supply Assembly, 120/240 VAC 50/60 Hz
 

80000-02300

Power Supply Assembly, 24 VDC
 

01902-17106 External Serial Output Cable

01902-30011 Keypad Assembly

24601-00002
Display Assembly
 
02203-13100
Daughter Board
 
45000-02002
Relay, DC12V SPDT 3 A/120 VAC
 

Herriott Cell (8m & 28m)
02101-17102
Pressure Transducer Assembly
 
02101-17103
Thermistor Assembly
 
02199-00006
Spares Kit (o-rings, fuses), AC, Viton
 
02199-00007
Cleaning Kit (domestic only)
 
 

Photos

 
 

« Previous | Next »

Request a quote

To request a quote online, please enter the following and we'll get back to you promptly:

Your Name*
Company
Email*
Address*
Address Line 2
City*
State
Zip Code
Country*
Phone
Product Code
Application Name / Stream Details
Measurement Range
Comments and Questions

« Previous | Next »