Component |
Unit |
Typical Concentration |
Min. for Application |
Max. for Application |
|
Acetylene (C2H2) |
ppmv |
1500 - 5000 |
0 |
5000 |
|
Ethylene (C2H4) |
Mole% |
65 |
60 |
90 |
|
Ethane (C2H6) |
Mole% |
33 |
0 |
40 |
|
Hydrogen (H2) |
ppmv |
0 |
0
|
1000
|
|
Carbon Monoxide (CO) |
ppmv |
0.5 |
0
|
1000 |
|
Carbon Dioxide (CO2) |
ppmv |
< 1.0 |
0 |
1000 |
|
Methane (CH4) |
ppmv |
50 - 100 |
0 |
1000 |
|
Propylene (C2H6) |
ppmv |
3000 |
0 |
5000 |
|
Total |
Mole% |
100 |
|
|
The background stream composition must be specified for proper calibration and measurement performance. Specify the Normal composition, along with the minimum and maximum expected values for each component, especially water, the measured component. Other stream compositions may be allowable with approval from SpectraSensors.
Performance |
|
|
Target Components |
Acetylene in Back End Acetylene Converter Inlet |
|
Typical Measurement Ranges |
0 – 5000 ppm (or greater)* |
|
Typical Precision |
±2% of Full Scale* |
|
Measurement Response Time |
1 to ~60 sec |
|
Principle of Measurement |
Tunable Diode Laser Absorption Spectroscopy
Non-differential |
|
Environmental Temperature Range |
-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional
|
|
Sample Inlet Pressure |
70 kPag (10 psig) typical
210 kPag (30 psig) maximum
|
|
Cell Temperature |
Maintain at 50ºC +/- 0.2ºC |
|
Maximum Cell Pressure |
70 kPag (10 PSIg) |
|
Sample Flow Rate
|
1 - 2 L/min (2.1 to 4.2 scfh)* |
| Validation |
Certified blend of C2H2 in Nitrogen balance |
Electrical |
|
|
Power |
100-240 VAC, 50-60 Hz standard |
|
Max Current |
Controller: 1 A @ 120 VAC |
|
Controller to Cell Cable Length |
1m standard (3m, 5m & 10m available optionally) |
|
Communication |
Current Loop Output 4-20 mA Isolated, 1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
|
| Digital Outputs |
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm
|
| LCD Display |
Concentration, Cell Pressure and Temperature, Diagnostic Data |
Physical
|
|
|
Controller Enclosure |
NEMA 4X - 304 Stainless Steel standard |
|
Controller Dimensions |
343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*
|
|
Weight Approximately |
13.1 Kg (28.6 lbs)* |
|
Sample Cell Dimensions |
0.8m cell
438 mm H × 108 mm W (17-1/4” H × 4-1/4” W)* |
|
Sample Cell Construction |
316L Series Polished Stainless Steel standard
|
| Number of Sample Cells |
1 (single channel SS2100) or
2 (dual channel SS2100)
|
| Area Classification |
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
|
| Dimensions with Sample System | 1678 mm H x 613 mm W x 427 mm D (66” H x 24-1/8” W x 16-13/16” D) |
| Weight with Sample System (approx.) | 68 kg (150 lbs.) |
|
* Application specific; consult factory
|
|
|
SS2100 ARM9 Electronics
|
|
|
80000-02199 |
Temperature Control Board
|
|
23000-00030 |
4-20 mA Current Loop Board
|
|
80000-02203 |
Power Supply Assembly, 120/240 VAC 50/60 Hz
|
|
80000-02300 |
Power Supply Assembly, 24 VDC
|
| 01902-17106 | External Serial Output Cable
|
| 01902-30011 | Keypad Assembly
|
| 24601-00002 |
Display Assembly
|
| 02203-13100 |
Daughter Board
|
| 45000-02002 |
Relay, DC12V SPDT 3 A/120 VAC
|
| Cell (0.8m) | |
| 01902-17102 |
Pressure Transducer Assembly
|
| 02199-00001 |
Spares/Cleaning Kit (o-rings, fuses), AC, Viton (domestic only)
|
| 02199-00005 |
Spares Kit (o-rings, fuses), AC, Viton
|