SS2100 - Dry Cracked Gas (Dryer Outlet) Analyzer Solution

In Ethylene production the process gas is cooled to approximately -150oC in the separation section of the plant where the various products will be cryogenically fractionated. Water must be removed prior to cooling the process gas. The drier the process gas is, the longer and more efficiently the separation section can be operated without shut-downs.
Product Code: 54801

 
 
 
 
 

Application Note

Cracked Gas Dryer
Molecular Sieves and Activated Alumina are both used in Cracked Gas drying. As the inlet gas is saturated, water concentration can be estimated from the process temperature and pressure, and the cycle time of the dryers is controlled from their known capacity. However, dryers are subject to fouling with polymers and heavy hydrocarbons and channeling through the dryer bed can shorten the effective dryer cycle time.

Critical Control of Moisture
To maintain the efficiency of the process and to ensure the purity of the final Ethylene product, water measurements at the outlet of the dryer are essential.

Traditional Measurement Solutions
Vibrating crystal, AlO3 capacitance probes and electrolytic moisture analyzers have been the traditional methods for monitoring the levels of water in Polymer Grade Ethylene. These methods suffer from slow wet-up and dry-down times, slow drift in calibration and sensitivity to contamination Due to the very short residence times in Ethylene feed storage vessels and the desired to measure Ethylene in flowing pipelines, the delay in measurements can easily lead to excursions in concentrations before it is detected. Due to their slow drift in calibration, capacitance probes have very high maintenance costs due to frequent replacement of the probe.

SpectraSensors' Solution
The SpectraSensors SS2100 is the ideal solution for moisture in Dry Cracked Gas. The use of Tunable Diode Laser technology means that analysis results can be updated every second. Furthermore, the high resolution that is inherent to the laser eliminates errors due to interferences that have hampered other spectrometric approaches.

Also, the non-contact nature of the measurement with no moving parts in the analyzer means that the analyzer is simple to operate and little routine maintenance is needed. Only SpectraSensors applies Differential Spectroscopy where the sample is dried using a metal getter technology and the background spectrum of the sample is collected.

The dryer is then bypassed and the wet sample spectrum is collected The dry spectrum is subtracted from the wet spectrum to get a sensitive measurement free of background gas interferences, and thus changing background gas compositions are compensated. The dryer has a typical lifetime greater than 18 months, and is the only consumable part of the analyzer. See Fig. 3 for typical calibration performance.


Typical Stream Composition for Dry Cracked Gas:

Component

Unit

Typical Concentration

Min. for Application

Max. for Application

Water

ppmv

< 5

0.05

10

Hydrogen Sulfide

(H2S)

ppmv

< 2

0

10

Carbon Dioxide (CO2)

ppmv

<5

0

10

Hydrogen (H2)

Mole%

25

10

 

30

 

Methane (CH4)

Mole%

20

10

35

Ethane (C2H6)

Mole%

15

5

30

Ethylene (C2H4)

Mole%

25

20

40

Acetylene (C2H2)

Mole%

0.3

0

0.5

Propylene (C2H6)

Mole%

7.5

0

15

Propane (C3H8)

Mole%

7.5

0

15

Methyl Acetylene

(Propyne C3H4)

Mole%

0.03

0

0.1

Propadiene (C3H4)

Mole%

0.02

0

0.1

Carbon Monoxide (CO)

Mole%

0.05

0

 

0.1

Butanes

Mole%

0.05

0

1.0

Butenes

Mole%

0.3

0

1.0

1,3-Butadiene

Mole%

0.5

0

2.0

C5+

Mole%

0.1

0

2.0

Total

Mole%

100

 

 

 

The background stream composition must be specified for proper calibration and measurement performance. Specify the Normal composition, along with the minimum and maximum expected values for each component, especially water, the measured component. Other stream compositions may be allowable with approval from SpectraSensors. 


 
 
 

« Previous | Next »

Data Sheet


Performance

Target Components

H2O in Dry Cracked Gas

Typical Measurement Ranges

0 – 10ppm* (other ranges available by request)

Typical Precision

 ±2% of Full Scale*

Measurement Response Time

1 to ~60 sec

Principle of Measurement

Differential Tunable Diode Laser Absorption Spectroscopy
(H2O Scrubber included)

Environmental Temperature Range

-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional

Sample Inlet Pressure

70 kPag (10 psig) typical

210 kPag (30 psig) maximum


Cell Temperature

Maintain at 50ºC +/- 0.2ºC

Maximum Cell Pressure

70 kPag (10 PSIg)

Sample Flow Rate
3 - 4 L/min (6.4 to 8.5 scfh)*
Validation
A certified blend of H2O in Nitrogen balance is diluted
with dried sample in the sample conditioning system
under flow control
 

Electrical

Power

100-240 VAC, 50-60 Hz standard

Max Current

Controller: 1 A @ 120 VAC

Controller to Cell Cable Length

1m standard (3m, 5m & 10m available optionally)

Communication

Current Loop Output 4-20 mA Isolated,
1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
Digital Outputs
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm 
LCD Display
Concentration, Cell Pressure and Temperature, Diagnostic Data
 

Physical

Controller Enclosure

NEMA 4X - 304 Stainless Steel standard

Controller Dimensions

343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*

Weight Approximately

13.1 Kg (28.6 lbs)*

Sample Cell Dimensions

28m Herriot Cell
559 mm H × 127 mm W
(22” H × 5” W)

Sample Cell Construction

316L Series Polished Stainless Steel standard
SilcoNert® coated
Number of Sample Cells
1 (single channel SS2100) or
2 (dual channel SS2100)
Area Classification
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
 ll 2G Ex d llB+H2 T5; Tamb : -20 ÷ +60 °C
Dimensions with Sample System 1678 mm H x 613 mm W x 427 mm D
(66” H x 24-1/8” W x 16-13/16” D)
Weight with Sample System (approx.) 68 kg (150 lbs.)

* Application specific; consult factory

« Previous | Next »

Drawings


 

Typical Accessories


SS2100 ARM9 Electronics
 

80000-02199

Temperature Control Board

23000-00030

4-20 mA Current Loop Board
 

80000-02203

Power Supply Assembly, 120/240 VAC 50/60 Hz
 

80000-02300

Power Supply Assembly, 24 VDC
 

01902-17106 External Serial Output Cable

01902-30011 Keypad Assembly

24601-00002
Display Assembly
 
02203-13100
Daughter Board
 
45000-02002
Relay, DC12V SPDT 3 A/120 VAC
 

Herriott Cell (8m & 28m)
02101-17102
Pressure Transducer Assembly
 
02101-17103
Thermistor Assembly
 
02199-00006
Spares Kit (o-rings, fuses), AC, Viton
 
02199-00007
Cleaning Kit (domestic only)
 
 
 

Photos

 
 

« Previous | Next »

Request a quote

To request a quote online, please enter the following and we'll get back to you promptly:

Your Name*
Company
Email*
Address*
Address Line 2
City*
State
Zip Code
Country*
Phone
Product Code
Application Name / Stream Details
Measurement Range
Comments and Questions

« Previous | Next »