SpectraSensors OXY4400 Oxygen Analyzer is a compact stand-alone one-channel meter with LCD display and data logger. The electronics are certified for hazardous area use. The sensor probe is inserted into the process stream and is connected to the controller by an optical fiber.
Full sample conditioning systems are available with the OXY4400 as well as software for PC interface.
Product Code: 11810
Application Note
Low
Maintenance The analyzer uses an optical method that detects oxygen using a probe that is inserted into the gas stream. The probe can be easily cleaned and has a lifetime measured in years.
Calibration of the analyzer is a simple procedure that can be performed in minutes using a binary standard with oxygen in nitrogen.
Accurate and ReliableThe OXY4400 technology is ideally suited for measuring in Natural Gas.
The sensor is not affected by even high levels of H2S or other sulfur species. There is no cross sensitivity to contaminants or other gases in natural gas.
The accuracy and reliability of the measurements are superior to electrochemical analyzers because the measurement does not drift.
« Previous | Next »
Data Sheet
Performance |
|
Target Components
|
O2
|
|
Typical Measurement Ranges
|
0-100 ppmv default, 0-10 to 0-1,000 ppmv user settable |
|
Lower Limit of Detection |
0.5 ppmv |
|
Accuracy at 20°C |
±5% of Reading (10±0.5 ppmv, 100±5 ppmv) |
|
Measurement Response Time |
Programmable Sampling Rate (Default 3 Second) |
|
Principle of Measurement |
Fluorescent Quenching
|
|
Environmental Temperature Range |
0 to 50ºC (32 to 122ºF)
|
|
Sample Inlet Pressure |
140-350 kPaG (20-50 PSIG) to sample panel regulator |
|
Sample Probe Temperature Range |
Up to 50ºC |
|
Maximum Probe Pressure
|
200 psig |
| Sample Flow Rate |
1.5 slpm (0.05 scfm) |
| Recommended Calibration |
Conventional two-point calibration in oxygen-free
environment (nitrogen) and a second point
100 ppm (300 ppm Maximum) |
| |
Electrical & Communications
|
|
Input Power |
100/240 VAC 50/60 Hz and 9/24 VDC max 0.83 Amps at 24 VDC |
|
Controller to Probe Cable Length |
0.7m default (2.5m optional) |
|
Communication |
Analog: Qty 2 4-20 mA and Qty 2 0-10V,
Independent, Isolated
Serial: RS232C Interface, Isolated, 19200 Baud,
8 data-bits |
|
Serial |
RS232C Interface, Isolated, 19200 Baud, 8 data-bits
|
| |
Physical
|
|
Enclosure Type |
NEMA Type 4 |
|
Analyzer Dimensions |
203 X 254 X 203mm (8 X 10 X 8 inches)
|
|
Weight |
2.2 kg (4.9 lb) |
|
Sample Probe Construction |
316 Stainless Steel |
|
Number of Measurement Channels |
1 |
Certification
|
Class I, Division II, Groups A, B, C, and D ID 3035281 & 3035281C (CSA C22.2 No. 142 & 213) |
|
|
« Previous | Next »